GE AMX 4+

Sold

GE AMX 4+

Description

GE AMX4+ Portable X-ray

  • Analog X-ray system in a compact, easy-to-maneuver package.
  • Up to 50 high-quality exposures with a single charge.
  • Dual drive motors and oversized casters that facilitate movement even in taxing environments.
  • Operator controls are microprocessor-based, combining the latest technology with simplicity of design.
  • Message display to provide helpful information on the operating status of the system.
  • Ergonomic Handswitch Controls.
  • Compact manual Collimator with built-in SIDbutton on collimator face.

Dimensions

  • Height: 70 inches (177.8 cm) for models 2169360−6, 2236420−6, and 2275938−6, −12, −13, −14, −15
  • Height: 76 inches (193 cm) for all other models.
  • Width: 25−3/16 inches (64 cm)
  • Length: 45−3/8 inches (115.3 cm)
  • Weight: 1080 pounds (490 kg)

Movements

  • Tube vertical movement measured at the focal spot (arm extended)
  • Range at least 46.5 inches
  • Lowest position 26.1 inches
  • Highest position 72.6 inches)
  • The horizontal movement measured at the focal spot relative to column face is 24 inches (61.0 cm) minimum, to 40 inches (101.6 cm) maximum.
  • Tube Column rotation measured from horizontal arm latch is +/−270 degrees.
  • Tube and yoke rotation around Horizontal Arm measured from tube port down position:
  • Range 360 degrees;
  • Detent locations, +/−90, and +/−180 degrees.
  • Tube Trunnion rotation measured from tube port down position:
  • Range 120 degrees;
  • Forward 110 degrees;
  • Backward 10 degrees;
  • Detent 0 degrees, and 90 degrees.
  • Collimator Rotation measured from the front of the collimator with the tube port facing down:
  • Range 180 degrees;
  • Right 90 degrees;
  • Left 90 degrees;
  • Detent 0 and 9

Drive Speed

  • Drive Speed is 264 feet (670.5 cm) per minute +/− 25%;
  • Maneuvering Speed is 30% to 60% of drive speed.

kVp Accuracy

  • Rise time of the kVp wave form from 10% to 90% of the maximum kVp is 1.2 millisecond or less.
  • Fall time of the kVp wave form from 90% of the maximum kV to 20 kV is 2.5 milliseconds or less.
  • Accuracy of the kVp wave form to selected kVp is +/− 8% of the value displayed on the operator panel for the first 20 ms and +/− 5% after 20 ms. Accuracy applies within the range of the bar graph battery charge indicator.

Collimator Function

  • Type: Manual
  • Minimum source to skin distance is limited to more than 30 centimeters by the skin spacers at the beam exit of the collimator.
  • Full 17 by 17 inch (43 by 43 centimeters) radiographic coverage at 40 inch (1.02 meter) Source to Image Distance.
  • Minimum inherent filtration of 2.0 mm aluminum equivalent at 100 kVp.

Collimator Light Field Intensity

  • The average illumination at a distance of 100 cm (39.37 inch) from the focal spot shall be 16 foot candles (170 lux) or more.

Generator Performance and Tubes

  • 0.75 mm focal spot (NEMA): 3-inch rotating anode: GE X-Ray tube model HRT09.
  • 275,000 HU anode heat storage capacity.
  • 15° target angle.
  • Low speed (3,000 rpm) operation only (See Product Data Sheet D1046)
  • The generator is closed loop kVp design using microprocessor regulation to assure constant and accurate kVp at all battery conditions.

Battery Specifications

  • Nine 12.9 volt batteries connected in series provide approximately 116 volts at full charge.

Battery Capacity

  • The GE AMX 4+ battery capacity can be measured by one of the following five methods. All capacities are measured after the AMX−4+ has been charged to the “CHARGE COMPLETE” state. Available capacity as stated applies only to new battery sets free of defective cells. Capacity may decrease as the battery nears the end of its useful life


Specifications

ManufacturerGE
ModelAMX 4+
ConditionUsed